Abstract: | In this study, a new analytical method for erdosteine (ERD) in plasma based on high‐performance liquid chromatography and a fluorimetric detector, is presented. Precolumn derivatization of ERD with 4‐bromomethyl‐7‐methoxy coumarin (BrMmC) and dibenzo‐18‐crown‐6‐ether as a reaction catalyst led to the production of a fluorescent compound. ERD was monitored by fluorescence with an excitation wavelength λext. = 325 nm and emission wavelength λem. = 390 nm. Optimum reaction conditions were carefully studied and optimized. A chromatographic procedure was performed using a C18 column of 150 × 4.6 mm and 3 μm particle size and a mobile phase consisting of methanol:acetonitrile:water (30:30:40, v/v/v) under a flow rate of 0.5 ml min?1. A calibration plot was established covering analyte concentration range 0.2–3.0 μg ml?1; the detection limit was 0.015 μg ml?1 and quantification limit was 0.05 μg ml?1. Mean recovery was 87.33% and relative standard deviation was calculated to be less than 4.4%. The developed method was successfully used to determine pharmacokinetic preparations of ERD subsequent to administration of a 900 mg dose capsule to a healthy 40‐year‐old woman volunteer. |