首页 | 本学科首页   官方微博 | 高级检索  
     


Oxygen tolerance of an implantable polymer/enzyme composite glutamate biosensor displaying polycation-enhanced substrate sensitivity
Authors:McMahon Colm P  Rocchitta Gaia  Kirwan Sarah M  Killoran Sarah J  Serra Pier A  Lowry John P  O'Neill Robert D
Affiliation:UCD School of Chemistry and Chemical Biology, University College Dublin, Belfield, Dublin 4, Ireland.
Abstract:Biosensors were fabricated at neutral pH by sequentially depositing the polycation polyethyleneimine (PEI), the stereoselective enzyme L-glutamate oxidase (GluOx) and the permselective barrier poly-ortho-phenylenediamine (PPD) onto 125-microm diameter Pt wire electrodes (Pt/PEI/GluOx/PPD). These devices were calibrated amperometrically at 0.7 V versus SCE to determine the Michaelis-Menten parameters for enzyme substrate, l-glutamate (Glu) and co-substrate, dioxygen. The presence of PEI produced a 10-fold enhancement in the detection limit for Glu (approximately 20 nM) compared with the corresponding PEI-free configurations (Pt/GluOx/PPD), without undermining their fast response time (approximately 2 s). Most remarkable was the finding that, although some designs of PEI-containing biosensors showed a 10-fold increase in linear region sensitivity to Glu, their oxygen dependence remained low.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏