首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Development of photocatalytic biosensor for the evaluation of biochemical oxygen demand
Authors:Chee Gab-Joo  Nomura Yoko  Ikebukuro Kazunori  Karube Isao
Institution:Research Center for Advanced Science and Technology, The University of Tokyo, 4-6-1 Komaba Meguro-ku, Tokyo 153-8904, Japan. cheegj@jamstec.go.jp
Abstract:The photocatalytic biosensor of flow system using semiconductor TiO2 was developed to evaluate biochemical oxygen demand (BOD) levels in river water. Photocatalysis of sample was carried out in a photoreactor with TiO2 and a 6W black-light blue fluorescent tube as light source. Sample from a photoreactor outlet was measured by an oxygen electrode with a biofilm. The sensor response of photocatalytic biosensor was between 5 and 10 min depending on concentration of biochemical in the samples. At BOD of 1 mgl-1, the sensor response increased 1.33-fold in comparison with that without photocatalysis. The degradation of tannic acid and humic acid with photocatalysis were 51.8 and 38.4%, respectively. Gum arabic and linear alkylbenzene sulfonate (LAS) were degraded a little, but gave the responses of more than double to the sensor. Free radicals yielded by photocatalysis in a photoreactor did not affect the sensor response because their lifetime is extremely short. Fairly good correlation (r=0.983) between the sensor method and the conventional method was obtained for test samples. This biosensor using photocatalytic pretreatment improved the sensitivity.
Keywords:
本文献已被 PubMed 等数据库收录!