首页 | 本学科首页   官方微博 | 高级检索  
     


In vitro assessment of the influence of aortic annulus ovality on the hydrodynamic performance of self-expanding transcatheter heart valve prostheses
Authors:Maximilian Kuetting  Alexander Sedaghat  Marc Utzenrath  Jan-Malte Sinning  Christoph Schmitz  Jan Roggenkamp  Nikos Werner  Thomas Schmitz-Rode  Ulrich Steinseifer
Affiliation:1. Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Helmholtz Institute, RWTH Aachen, University, Aachen, Germany;2. Med. Klinik und Poliklinik II, Universitätsklinikum Bonn, Rheinische Friedrich-Wilhelms-Universität Bonn, Germany
Abstract:

Background

Although CT-studies as well as intraoperative analyses have described broad anatomic variations of the aortic annulus, which is predominantly found non-circular, commercially available transcatheter aortic heart valve prostheses are circular. In this study, we hypothesize that the in vitro hydrodynamic function of a self-expanding transcatheter heart valve (Medtronic CoreValve®) assessed in an oval compartment representing the aortic annulus will differ from the conventionally used circular compartment.

Methods

Medtronic CoreValve® prostheses were tested in specifically designed and fabricated silicone compartments with three degrees of defined ovalities. The measurements were performed in a left heart simulator at three different flow rates. In this setting, regurgitation flow, effective orifice area, and systolic pressure gradient across the valve were determined. In addition, high speed video recordings were taken to investigate leaflet kinematics.

Results

The pressure difference across the prosthesis increased with rising ovality. The effective orifice areas were only slightly impacted. The analyses of the regurgitation showed minor changes and partially lower regurgitation when switching from round to slightly oval settings, followed by strong increases for further ovalization. The high speed videos show minor central leakage and impaired leaflet apposition for strong ovalities, but no leaflet/stentframe contact in any setting.

Conclusion

This study quantifies the influence of oval expansion of transcatheter heart valve prostheses on their hydrodynamic performance. While slight ovalities were well tolerated by a self-expanding prosthesis, more significant ovality led to worsening of prosthesis function and regurgitation.
Keywords:Aortic valve stenosis   Aortic valve prosthesis   Transcatheter aortic valve replacement   TAVI
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号