心室纤颤信号的自适应检测
引用本文:朱贻盛,李丹,王妙发.心室纤颤信号的自适应检测[J].生物物理学报,1994,10(1):149-153.
作者姓名:朱贻盛  李丹  王妙发
作者单位:上海交通大学仪器工程系生物医学工程和仪器实验室,美国麻省理工学院海洋工程系
摘    要:在分析了心室纤颤信号和心动过速信号自相关性的基础上,运用自适应LMS算法。以误差信号能量作为特征值,成功地检测了心室纤颤信号,比仅依赖心率来判别心跳是否正常可靠性要高一些。通过对156个数据学习和三次判断,其心室纤颤信号识别率达到97.4%,心动过速信号识别率达到100%.

关 键 词:自相关性,自适应LMS算法,误差信号能量

ADAPTIVE DETECTION FOR VENTRICULAR FIBRILIATION
Zhu Yisheng ,Li Dan ,Wang Miaofa.ADAPTIVE DETECTION FOR VENTRICULAR FIBRILIATION[J].Acta Biophysica Sinica,1994,10(1):149-153.
Authors:Zhu Yisheng  Li Dan  Wang Miaofa
Abstract:With the energy of the error signal (EE) as its characteristic value, Ventricular Fibrillation(VF) was successfully using adaphve LMS algorithm bed on the analysis ofautocowhtion of VF and VT (Ventricular Tachycaldin) signal. The reliability of this approachis higher than what is determined only by counting the ventricular rate. Through the learningof 156 episodes and judgements, correct identification rate of VF and VT arched 97.4%and 100% respectively.
Keywords:Autoconelation Adaphve LMS algorithm The energy of the error signal  