首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Cobalt(II) salophen-modified carbon-paste electrode for potentiometric and voltammetric determination of cysteine
Authors:Amini Mohammad K  Khorasani Jafar H  Khaloo Shokooh S  Tangestaninejad Shahram
Institution:Chemistry Department, University of Isfahan, Isfahan 81746-73441, Iran. mkamini@sci.ui.ac.kr
Abstract:A chemically modified electrode constructed by incorporating N,N(')-bis(salicylidene)-1,2-phenylenediaminocobalt(II) into carbon-paste matrix was used as a sensitive electrochemical sensor for detection of cysteine. The resulting electrode exhibits catalytic properties for the electrooxidation of cysteine and lowers the overpotential for the oxidation of this compound. The faster rate of electron transfer results in a near-Nernstian behavior of the modified electrode and makes it a suitable potentiometric and voltammetric sensor for the fast and easy determination of cysteine. A linear response in concentration range from approximately 2 microM to 0.01 M was obtained with a detection limit of 1 microM for the potentiometric detection of cysteine. The modified electrode was also used for the amperometric and differential pulse voltammetric determination of cysteine and the results were compared with those of the potentiometric method.
Keywords:Modified electrode  Sensors  Carbon-paste electrode  Potentiometric sensor  Cysteine  Cobalt(II) salophen
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号