Rapid determination of plasmid-carrying yeast cells by using an imaging sensor system
Authors:Endo H  Suzuki M  Sode K  Tamiya E  Karube I
Institution:Research Center for advanced Science and Technology, University of Tokyo, 4-6-1 Komaba, Meguro-ku, Tokyo 153 Japan.
Abstract:A novel imaging sensor system for the determination of plasmid carrying yeast cells was developed. The sensor system consisted of an Silicon Intensifier Target (SIT) video camera, a fluorescent microscope, and a personal computer system equipped with an image memory board. This system was based on the fact that the membrane integrity of only plasmid-carrying cells is lost following cell growth in 5-fluoro-orotic acid (5-FOA) containing medium, and consequently these target cell can be stained with fluorescent probes and detected. In this study, plasmid-carrying cells were detected and their fraction determined in a mixture of both plasmid-carring and plasmid-free cells. A good correlation was observed between the values determined by this sensor system and the conventional method in the 30%-80% range, and one assay was possible within 4 h. This sensor system could be used for the monitoring of plasmid-carrying fraction in recombinant yeast cells during cultivation.
Keywords:plasmid  yeast  detection  sensor  image  analysis  5-fluoro-orotic acid  determination