Accuracy of Continuous Glucose Monitoring Measurements in Normo-Glycemic Individuals
Authors:Abimbola A. Akintola  Raymond Noordam  Steffy W. Jansen  Anton J. de Craen  Bart E. Ballieux  Christa M. Cobbaert  Simon P. Mooijaart  Hanno Pijl  Rudi G. Westendorp  Diana van Heemst
Affiliation:1 Department of Gerontology and Geriatrics, Leiden University Medical Center, Leiden, the Netherlands, ; 2 Department of Clinical Chemistry, Leiden University Medical Center, Leiden, The Netherlands, ; 3 Department of Endocrinology, Leiden University Medical Center, Leiden, the Netherlands, ; 4 Department of Public Health, University of Copenhagen, Copenhagen, Denmark, ; University of Catanzaro Magna Graecia, ITALY,
Abstract:BackgroundThe validity of continuous glucose monitoring (CGM) is well established in diabetic patients. CGM is also increasingly used for research purposes in normo-glycemic individuals, but the CGM validity in such individuals is unknown. We studied the accuracy of CGM measurements in normo-glycemic individuals by comparing CGM-derived versus venous blood-derived glucose levels and measures of glycemia and glycemic variability.MethodsIn 34 healthy participants (mean age 65.7 years), glucose was simultaneously measured every 10 minutes, via both an Enlite® CGM sensor, and in venous blood sampled over a 24-hour period. Validity of CGM-derived individual glucose measurements, calculated measures of glycemia over daytime (09:00h-23:00h) and nighttime (23:00h-09:00h), and calculated measures of glycemic variability (e.g. 24h standard deviation [SD]) were assessed by Pearson correlation coefficients, mean absolute relative difference (MARD) and paired t-tests.ResultsThe median correlation coefficient between CGM and venous glucose measurements per participant was 0.68 (interquartile range: 0.40–0.78), and the MARD was 17.6% (SD = 17%). Compared with venous sampling, the calculated measure of glycemia during daytime was 0.22 mmol/L higher when derived from CGM, but no difference was observed during nighttime. Most measures of glycemic variability were lower with CGM than with venous blood sampling (e.g., 24h SD: 1.07 with CGM and 1.26 with venous blood; p-value = 0.004).ConclusionIn normo-glycemic individuals, CGM-derived glucose measurements had good agreement with venous glucose levels. However, the measure of glycemia was higher during the day and most measures of glycemic variability were lower when derived from CGM.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号