Development of immunosensors for direct detection of three wound infection biomarkers at point of care using electrochemical impedance spectroscopy
Authors:Ciani Ilenia  Schulze Holger  Corrigan Damion K  Henihan Grace  Giraud Gerard  Terry Jonathan G  Walton Anthony J  Pethig Ronald  Ghazal Peter  Crain Jason  Campbell Colin J  Bachmann Till T  Mount Andrew R
Affiliation:School of Chemistry & EaStCHEM, The University of Edinburgh, Joseph Black Building, King's Buildings, West Mains Road, Edinburgh EH9 3JJ, UK.
Abstract:A method for label-free, electrochemical impedance immunosensing for the detection and quantification of three infection biomarkers in both buffer and directly in the defined model matrix of mock wound fluid is demonstrated. Triggering Receptor-1 Expressed on Myeloid cells (TREM-1) and Matrix MetalloPeptidase 9 (MMP-9) are detected via direct assay and N-3-oxo-dodecanoyl-l-HomoSerineLactone (HSL), relevant in bacterial quorum sensing, is detected using a competition assay. Detection is performed with gold screen-printed electrodes modified with a specific thiolated antibody. Detection is achieved in less than 1h straight from mock wound fluid without any extensive sample preparation steps. The limits of detection of 3.3 pM for TREM-1, 1.1 nM for MMP-9 and 1.4 nM for HSL are either near or below the threshold required to indicate infection. A relatively large dynamic range for sensor response is also found, consistent with interaction between neighbouring antibody-antigen complexes in the close-packed surface layer. Together, these three novel electrochemical immunosensors demonstrate viable multi-parameter sensing with the required sensitivity for rapid wound infection detection directly from a clinically relevant specimen.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号