首页 | 本学科首页   官方微博 | 高级检索  
   检索      


New dual electrochemical detector for microbore liquid chromatography determination of dopamine and serotonin in rat striatum dialysates
Authors:Fu-Chou Cheng  Ying Shih  Yea-Jiuan Liang  Lin-Lan Yang  Chung-Shi Yang
Institution:aDepartment of Medical Research, Taichung Veterans General Hospital, Taichung 402, Taiwan;bDepartment of Chemistry, National Chung-Hsing University, Taichung 402, Taiwan;cDepartment of Applied Cosmetology, Hung-Kuang Institute of Medical and Nursing Technology, Taichung 433, Taiwan
Abstract:A new type of liquid chromatographic (LC) dual thin-layer amperometric detector for the simultaneous measurement of trace levels of dopamine and serotonin in microdialysates is described. The concentrations of these analytes in rat dialysates are usually in the sub-nanomolar concentration range (typically, 0.10–5.00 pg in 5-μl dialysates). With this dual electrode, a glass-lined microbore column provides excellent sensitivity, selectivity, and separation. In addition, a three- to five-fold improvement in anodic current or cathodic responses over conventional dual electrodes in microbore LC can be achieved. Due to the irreversible electrochemical properties of some interference peaks, this dual electrode provides reliable measurement of dopamine based on the cathodic signal. The detection limit (signal-to-noise RATIO=3) of this assay is 0.02 pg per injection for dopamine or serotonin. This new dual electrode allows the simultaneous measurements of basal dopamine and serotonin in rat striatum dialysates without the use of re-uptake inhibitors in perfusion medium.
Keywords:Microdialysis  Dopamine  Serotonin  5-Hydroxytryptamine
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号