Direct quantification of human serum albumin in human blood serum without separation of gamma-globulin by the total internal reflected resonance light scattering of thorium-sodium dodecylbenzene sulfonate at water/tetrachloromethane interface
Authors:Feng Ping  Huang Cheng Zhi  Li Yuan Fang
Affiliation:Institute of Environmental Chemistry, Southwest Normal University, Chongqing 400715, People's Republic of China.
Abstract:A direct quantification of human serum albumin (HSA) in blood serum samples without separation is proposed based on the measurements of total internal reflected resonance light scattering (TIR-RLS) at water/tetrachloromethane (H(2)O/CCl(4)) interfaces. In the pH range of 6.37-6.59, the coadsorption of the binary complex of HSA-Th(IV) with sodium dodecylbenzene sulfonate occurs at the H(2)O/CCl(4) interface, forming an amphiphilic layer and displaying greatly enhanced TIR-RLS signals with the maximum peak located at 340-370 nm. The enhanced TIR-RLS intensity is in proportion to the HSA concentration in the range 0.15-1.0 micro gml(-1). The limit of detection is 14.4 ngml(-1). The contents of HSA in blood serum samples were determined with the recovery of 97.1-102.3% and RSD of 0.6-2.9%, which are identical to those obtained according to the spectrofluorimetric method using chrome azurol S.
Keywords:Total internal reflected resonance light scattering   Human serum albumin   Sodium dodecylbenzene sulfonate   Thorium (IV)