首页 | 本学科首页   官方微博 | 高级检索  
   检索      


A versatile shear and compression apparatus for mechanical stimulation of tissue culture explants
Authors:Frank E H  Jin M  Loening A M  Levenston M E  Grodzinsky A J
Institution:Continuum Electromechanics Group, Center for Biomedical Engineering, Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Room 38-377, Cambridge, MA 02139, USA. ehfrank@mit.edu
Abstract:We have developed an incubator housed, biaxial-tissue-loading device capable of applying axial deformations as small as 1 microm and sinusoidal rotations as small as 0.01 degrees. Axial resolution is 50 nm for applying sinewaves as low as 10 microm (or 1% based on a 1 mm thickness) or as large as 100 microm. Rotational resolution is 0.0005 degrees. The machine is small enough (30 cm high x 25 cm x 20 cm) to be placed in a standard incubator for long-term tissue culture loading studies. In metabolic studies described here, application of sinusoidal macroscopic shear deformation to articular cartilage explants resulted in a significant increase in the synthesis of proteoglycan and proteins (uptake of (35)S-sulfate and (3)H-proline) over controls held at the same static offset compression.