Flow-through immunofiltration assay system for rapid detection of E. coli O157:H7
Authors:Abdel-Hamid I  Ivnitski D  Atanasov P  Wilkins E
Institution:Department of Chemical and Nuclear Engineering, University of New Mexico, Albuquerque 87131, USA.
Abstract:A flow-through amperometric immunofiltration assay system based on disposable porous filter-membranes for rapid detection of Escherichia coli O157:H7 has been developed. The analytical system utilizes flow-through, immunofiltration and enzyme immunoassay techniques in conjunction with an amperometric sensor. The parameters affecting the immunoassay such as selection of appropriate filter membranes, membrane pore size, antibody binding capacity and the concentrations of immunoreagents were investigated and optimized. Non-specific adsorption of the enzyme conjugate was investigated and minimized. A sandwich scheme of immunoassay was employed and the immunofiltration system allows to specifically and directly detect E. coli cells with a lower detection limit of 100 cells/ml. The working range is from 100 to 600 cells/ml with an overall analysis time of 30 min. No pre-enrichment was needed. This immunosensor can be easily adapted for assay of other microorganisms and may be a basis for a new class of highly sensitive bioanalytical devices for rapid quantitative detection of bacteria.