Abstract: | A method for the detection of 8-hydroxydeoxyguanosine by high-performance capillary electrophoresis (HPCE) was developed. Separations were performed in an uncoated silica capillary (44 cm × 75 μm I.D.) with a P/ACE system with diode-array detector. The separation of purine deoxynucleosides and 8-hydroxydeoxyguanosine was optimized with regard to pH, temperature, applied potential and hydrodynamic injection time. Optimum conditions were 20 mM borate buffer (pH 9.5), 25°C, 25 kV, 20 s load and detection at 254 nm. This method allowed the detection of 8-hydroxydeoxyguanosine in the presence of a 105-fold higher amount of deoxyguanosine. Isolated nuclei from K562 human hematopoietic cells were treated with 15 mM hydrogen peroxide for 2 h. The nuclei were extensively dialyzed and DNA was isolated, enzymatically hydrolyzed to the deoxynucleosides and analyzed by HPCE. DNA from hydrogen peroxide treated nuclei had a 4-fold higher content of 8-hydroxydeoxyguanosine than untreated controls. HPCE analysis of 8-hydroxydeoxyguanosine is fast and simple. Furthermore, it requires a very small sample volume, which makes it useful for biomedical and clinical applications. |