Friction in metal-on-metal total disc arthroplasty: effect of ball radius |
| |
Authors: | Moghadas Parshia Mahomed Aziza Hukins David W L Shepherd Duncan E T |
| |
Affiliation: | School of Mechanical Engineering, University of Birmingham, Edgbaston, Birmingham, B15 2TT, UK. pxm384@bham.ac.uk |
| |
Abstract: | Total disc arthroplasty (TDA) can be used to replace a degenerated intervertebral disc in the spine. There are different designs of prosthetic discs, but one of the most common is a ball-and-socket combination. Contact between the bearing surfaces can result in high frictional torque, which can then result in wear and implant loosening. This study was designed to determine the effects of ball radius on friction. Generic models of metal-on-metal TDA were manufactured with ball radii of 10, 12, 14 and 16 mm, with a radial clearance of 0.015 mm. A simulator was used to test each sample in flexion-extension, lateral bending and axial rotation at frequencies of 0.25, 0.5, 0.75, 1, 1.25, 1.5, 1.75 and 2 Hz under loads of 50, 600, 1200 and 2000 N, in new born calf serum. Frictional torque was measured and Stribeck curves were plotted to illustrate the lubrication regime in each case. It was observed that implants with a smaller ball radius showed lower friction and showed boundary and mixed lubrication regimes, whereas implants with larger ball radius showed boundary lubrication only. This study suggests designing metal-on-metal TDAs with ball radius of 10 or 12 mm, in order to reduce wear and implant loosening. |
| |
Keywords: | |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|