首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Stress distribution on the cusps of a polyurethane trileaflet heart valve prosthesis in the closed position.
Authors:K B Chandran  S H Kim  G Han
Institution:Department of Biomedical and Mechanical Engineering, University of Iowa, Iowa City 52242.
Abstract:In this paper, a finite element analysis of the stress distribution on the cusps of a polyurethane trileaflet heart valve prosthesis in the closed position is presented. The geometry of the valve was modified from a relationship proposed by Ghista and Reul (J. Biomechanics 10, 313-324, 1977). The effects of variations in stent height, leaflet thickness and coaptation area on the stress distribution were also analyzed. Analyses were performed with both rigid and flexible stents for the trileaflet valve in order to delineate the effect of stent flexibility on the leaflet stress distribution. The results showed that regions of stress concentration were present near the commissural attachment similar to those predicted with the bioprostheses. The stresses on the leaflets were reduced by increasing the stent height with both rigid and flexible stents. Selectively increasing the leaflet thickness near the commissures and also increasing the coaptation area did not prove to reduce the leaflet stresses when the stent flexibility was taken into account. The possible effect of high stresses on the structural integrity of polyurethane leaflets and its relationship with calcification is yet to be investigated.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号