首页 | 本学科首页   官方微博 | 高级检索  
   检索      


High-performance liquid chromatographic determination of iothalamic acid in human plasma and urine
Authors:Flavio Gaspari  Laura Mainardi  Piero Ruggenenti  Giuseppe Remuzzi
Institution:

Mario Negri Institute for Pharmacological Research, Laboratory of Clinical Pharmacokinetics, Via Gavazzeni 11, 24100 Bergamo Italy

Abstract:A simple, rapid and sensitive method for the determination of iothalamic acid (IA) in both plasma and urine is reported. After extraction with ethyl acetate, IA was determined by strong anion-exchange high-performance liquid chromatography with ultraviolet detection at 254 nm. The lower limit of detection was 0.5 μg/ml. The average recovery was 73 and 57% from plasma and urine, respectively. Linearity was found over the investigated concentration range (up to 500 μg/ml for plasma and up to 10.0 mg/ml for urine). The reproducibility of the technique was good (coefficient of variation less than 6%) as was the precision and accuracy (coefficient of variation less than 2.5%). No interference from endogenous substances or any of the common drugs tested was found.
Keywords:
本文献已被 ScienceDirect 等数据库收录!