首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of MS-275, a novel histone deacetylase inhibitor, in human plasma by liquid chromatography-electrospray mass spectrometry
Authors:Hwang Kyunghwa  Acharya Milin R  Sausville Edward A  Zhai Suoping  Woo Eunhee W  Venitz Jürgen  Figg William D  Sparreboom Alex
Affiliation:Clinical Pharmacology Research Core, Medical Oncology Clinical Research Unit, Center for Cancer Research, National Cancer Institute, National Institutes of Health, 9000 Rockville Pike, Building 10, Room 5A01, Bethesda, MD 20892, USA.
Abstract:A rapid method was developed for the quantitative determination of the novel histone deacetylase inhibitor, MS-275, in human plasma. Calibration curves were constructed in the range of 1-100 ng/ml, and were analyzed using a weight factor proportional to the nominal concentration. Sample pretreatment involved a one-step protein precipitation with acetonitrile of 0.1 ml samples. The analysis was performed on a column (75 mm x 4.6 mm i.d.) packed with 3.5 microm Phenyl-SB material, using methanol-10mM ammonium formate (55:45 (v/v)) as the mobile phase. The column effluent was monitored by mass spectrometry with positive electrospray ionization. The values for precision and accuracy were always < or =5.58 and <11.4% relative error, respectively. The method was successfully applied to examine the pharmacokinetics of MS-275 in a cancer patient.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏