Evaluation of the Diagnostic Accuracy of CareStart G6PD Deficiency Rapid Diagnostic Test (RDT) in a Malaria Endemic Area in Ghana,Africa
Authors:Dennis Adu-Gyasi  Kwaku Poku Asante  Sam Newton  David Dosoo  Sabastina Amoako  George Adjei  Nicholas Amoako  Love Ankrah  Samuel Kofi Tchum  Emmanuel Mahama  Veronica Agyemang  Kingsley Kayan  Seth Owusu-Agyei
Affiliation:Kintampo Health Research Centre, P O Box 200, Kintampo, Brong Ahafo, Ghana.; Centro de Pesquisa Rene Rachou/Fundação Oswaldo Cruz (Fiocruz-Minas), BRAZIL,
Abstract:

Background

Glucose-6-phosphate dehydrogenase (G6PD) deficiency is the most widespread enzyme defect that can result in red cell breakdown under oxidative stress when exposed to certain medicines including antimalarials. We evaluated the diagnostic accuracy of CareStart G6PD deficiency Rapid Diagnostic Test (RDT) as a point-of-care tool for screening G6PD deficiency.

Methods

A cross-sectional study was conducted among 206 randomly selected and consented participants from a group with known G6PD deficiency status between February 2013 and June 2013. A maximum of 1.6ml of capillary blood samples were used for G6PD deficiency screening using CareStart G6PD RDT and Trinity qualitative with Trinity quantitative methods as the “gold standard”. Samples were also screened for the presence of malaria parasites. Data entry and analysis were done using Microsoft Access 2010 and Stata Software version 12. Kintampo Health Research Centre Institutional Ethics Committee granted ethical approval.

Results

The sensitivity (SE) and specificity (SP) of CareStart G6PD deficiency RDT was 100% and 72.1% compared to Trinity quantitative method respectively and was 98.9% and 96.2% compared to Trinity qualitative method. Malaria infection status had no significant (P=0.199) change on the performance of the G6PD RDT test kit compared to the “gold standard”.

Conclusions

The outcome of this study suggests that the diagnostic performance of the CareStart G6PD deficiency RDT kit was high and it is acceptable at determining the G6PD deficiency status in a high malaria endemic area in Ghana. The RDT kit presents as an attractive tool for point-of-care G6PD deficiency for rapid testing in areas with high temperatures and less expertise. The CareStart G6PD deficiency RDT kit could be used to screen malaria patients before administration of the fixed dose primaquine with artemisinin-based combination therapy.