首页 | 本学科首页   官方微博 | 高级检索  
   检索      


GPU-Accelerated Framework for Intracoronary Optical Coherence Tomography Imaging at the Push of a Button
Authors:Myounghee Han  Kyunghun Kim  Sun-Joo Jang  Han Saem Cho  Brett E Bouma  Wang-Yuhl Oh  Sukyoung Ryu
Abstract:Frequency domain optical coherence tomography (FD-OCT) has become one of the important clinical tools for intracoronary imaging to diagnose and monitor coronary artery disease, which has been one of the leading causes of death. To help more accurate diagnosis and monitoring of the disease, many researchers have recently worked on visualization of various coronary microscopic features including stent struts by constructing three-dimensional (3D) volumetric rendering from series of cross-sectional intracoronary FD-OCT images. In this paper, we present the first, to our knowledge, "push-of-a-button" graphics processing unit (GPU)-accelerated framework for intracoronary OCT imaging. Our framework visualizes 3D microstructures of the vessel wall with stent struts from raw binary OCT data acquired by the system digitizer as one seamless process. The framework reports the state-of-the-art performance; from raw OCT data, it takes 4.7 seconds to provide 3D visualization of a 5-cm-long coronary artery (of size 1600 samples x 1024 A-lines x 260 frames) with stent struts and detection of malapposition automatically at the single push of a button.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号