Temporal resolution measurement of 128-slice dual source and 320-row area detector computed tomography scanners in helical acquisition mode using the impulse method
Affiliation:1. Department of Aquatic Life and Medical Sciences, Sun Moon University, Asan-si, Chungnam 336-708, South Korea;2. Department of Aquatic Life Medicine, Pukyong National University, Busan 608-737, South Korea
Abstract:PurposeTo analyse the temporal resolution (TR) of modern computed tomography (CT) scanners using the impulse method, and assess the actual maximum TR at respective helical acquisition modes.MethodsTo assess the actual TR of helical acquisition modes of a 128-slice dual source CT (DSCT) scanner and a 320-row area detector CT (ADCT) scanner, we assessed the TRs of various acquisition combinations of a pitch factor (P) and gantry rotation time (R).ResultsThe TR of the helical acquisition modes for the 128-slice DSCT scanner continuously improved with a shorter gantry rotation time and greater pitch factor. However, for the 320-row ADCT scanner, the TR with a pitch factor of <1.0 was almost equal to the gantry rotation time, whereas with pitch factor of >1.0, it was approximately one half of the gantry rotation time. The maximum TR values of single- and dual-source helical acquisition modes for the 128-slice DSCT scanner were 0.138 (R/P = 0.285/1.5) and 0.074 s (R/P = 0.285/3.2), and the maximum TR values of the 64 × 0.5- and 160 × 0.5-mm detector configurations of the helical acquisition modes for the 320-row ADCT scanner were 0.120 (R/P = 0.275/1.375) and 0.195 s (R/P = 0.3/0.6), respectively.ConclusionBecause the TR of a CT scanner is not accurately depicted in the specifications of the individual scanner, appropriate acquisition conditions should be determined based on the actual TR measurement.
Keywords:Temporal resolution  Helical acquisition mode  Gantry rotation time  Pitch factor
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号