Use of a control film piece in radiochromic film dosimetry
Institution:1. Radiation Physics Section, Biomedical Physics Department, King Faisal Specialist Hospital & Research Centre, Riyadh, Saudi Arabia;2. Medical Physics and Biophysics Division, Department of Radiation Oncology, Dana-Farber/Brigham Women''s Cancer Center, Harvard Medical School, Boston, MA, USA;3. Medical Physics Unit, McGill University, Montréal, Québec, Canada;4. Department of Radiation Oncology, SMBD Jewish General Hospital, Montréal, Québec, Canada;2. Imam Khomeini Hospital, Tehran University of Medical Sciences, Tehran, Iran;3. Iran Gamma Knife Center, Tehran, Iran;1. Riverina Cancer Care Centre, Wagga Wagga, New South Wales 2650, Australia;2. Radiation Oncology Queensland, Rockville, Queensland 4350, Australia;3. Epworth Radiation Oncology, Epworth HealthCare, Richmond, Victoria 3121, Australia
Abstract:PurposeRadiochromic films change their color upon irradiation due to polymerization of the sensitive component embedded within the sensitive layer. However, agents, other than monitored radiation, can lead to a change in the color of the sensitive layer (temperature, humidity, UV light) that can be considered as a background signal and can be removed from the actual measurement by using a control film piece. In this work, we investigate the impact of the use of control film pieces on both accuracy and uncertainty of dose measured using radiochromic film based reference dosimetry protocol.MethodsWe irradiated “control” film pieces (EBT3 GafChromicTM film model) to known doses in a range of 0.05–1 Gy, and five film pieces of the same size to 2, 5, 10, 15 and 20 Gy, considered to be “unknown” doses. Depending on a dose range, two approaches to incorporating control film piece were investigated: signal and dose corrected method.ResultsFor dose values greater than 10 Gy, the increase in accuracy of 3% led to uncertainty loss of 5% by using dose corrected approach. At lower doses and signals of the order of 5%, we observed an increase in accuracy of 10% with a loss of uncertainty lower than 1% by using the corrected signal approach.ConclusionsIncorporation of the signal registered by the control film piece into dose measurement analysis should be a judgment call of the user based on a tradeoff between deemed accuracy and acceptable uncertainty for a given dose measurement.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号