Abstract:A new liquid–liquid extraction is described for thiopurine methyl transferase (TPMT, EC 2.1.1.67) activity determination: the use of a pH 9.5 NH4Cl buffer solution, before adding the solvent mixture, allows more rapid extraction, avoiding a centrifugation step, and reduces the global cost of analysis. After the extraction step, 6-methylmercaptopurine, synthesised during the enzymatic reaction, is determined by a liquid chromatographic assay. Analytical performance of the assay was tested on spiked erythrocyte lysates. The linear concentration range was 5–250 ng ml −1 (r≥0.997, slope=1.497, intercept=−0.367). The recoveries were 82.8, 89.9 and 82.2% for 75, 125 and 225 ng ml−1, respectively. The coefficients of variation were ≤6.1% for within-day assay (n=6) and ≤9.5% for between-day assay precision (n=6; 14 days). TPMT activity was determined in a French adult Caucasian population (n=70). The results ranged from 7.8 to 27.8 nmol h−1 ml−1 packed red blood cells and the frequency distribution histogram is similar to that previously published.