Comparison of two techniques for quantifying environmental contaminants in human serum
Authors:J F Donahue  V W Burse  S L Head  J S Andrews
Institution:Division of Environmental Health Laboratory Sciences, Centers for Disease Control, Atlanta, GA 30333.
Abstract:Peak area matching and linear regression were used to quantify eight chlorinated pesticides and polychlorinated biphenyls (as Aroclor 1260) in human serum. There are no statistically significant differences in data obtained by these two quantifying techniques which were indicated by the paired t-test. For chlorinated pesticides, p = 0.053-0.62, and for polychlorinated biphenyls (as Aroclor 1260), p = 0.64. Analyte residues for the chlorinated pesticides ranged from 0.5 ppb for hexachlorobenzene (HCB) to 186 ppb for dichlorodiphenyldichloroethylene (DDE). Analyte residues for the polychlorinated biphenyls (as Aroclor 1260) ranged from 5-114 ppb. The absolute mean percent difference between the two quantifying techniques ranged from 0.06% for DDE to 8.06% for dieldrin (HEOD) among the chlorinated pesticides. The absolute mean percent difference between the two quantifying techniques for the polychlorinated biphenyls (as Aroclor 1260) was 3.4%. Peak area matching and linear regression were found to be comparable for quantifying these environmental residues in serum when the following conditions apply: 1) the concentration of the chlorinated pesticides is greater than or equal to 0.5 ppb (e.g., HCB, hexachlorocyclohexane (HCCH), oxychlordane (OC), heptachlor epoxide (HE), transnonachlor (TN), HEOD, and dichlorodiphenyltrichloroethane (DDT); 2) the concentration of the chlorinated pesticide is greater than or equal to 3 ppb (e.g., DDE); and 3) the total concentration of polychlorinated biphenyls (e.g., as Aroclor 1260) is greater than or equal to 5 ppb.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号