首页 | 本学科首页   官方微博 | 高级检索  
   检索      


A comparison of imaging methods for use in an array biosensor
Authors:Golden Joel P  Ligler Frances S
Institution:Naval Research Laboratory, Center for Bio/Molecular Science and Engineering, Code 6900, Washington, DC 20375-5348, USA. jgolden@cbmse.nrl.navy.mil
Abstract:An array biosensor has been developed which uses an actively-cooled, charge-coupled device (CCD) imager. In an effort to save money and space, a complementary metal-oxide semiconductor (CMOS) camera and photodiode were tested as replacements for the cooled CCD imager. Different concentrations of CY5 fluorescent dye in glycerol were imaged using the three different detection systems with the same imaging optics. Signal discrimination above noise was compared for each of the three systems.
Keywords:
本文献已被 PubMed 等数据库收录!