Feasibility of Performing Multiple Point of Care Testing for HIV Anti-Retroviral Treatment Initiation and Monitoring from Multiple or Single Fingersticks
Authors:Natasha Gous  Lesley Scott  Joachim Potgieter  Lumka Ntabeni  Sharon Enslin  Ronel Newman  Wendy Stevens
Institution:1. Department of Molecular Medicine and Haematology, School of Pathology, Faculty of Health Sciences, University of Witwatersrand, Johannesburg, South Africa.; 2. Department of Haematology, Tshwane Academic Division and University of Pretoria, Pretoria, South Africa.; 3. National Health Laboratory Service, Johannesburg, South Africa.; Fundacion Huesped, Argentina,
Abstract:

Background

Point of Care testing (POCT) provides on-site, rapid, accessible results. With current South African anti-retroviral treatment guidelines, up to 4 fingersticks /patient/clinic visit could be required if utilizing POC. We determined the feasibility and accuracy of a nurse performing multiple POCT on multiple fingersticks followed by simplification of the process by performance of multiple POC on a single fingerstick.

Method and Findings

Random HIV positive adult patients presenting at a HIV treatment clinic in South Africa, for ART initiation/ monitoring, were approached to participate in the study between April-June 2012. Phase I: n=150 patients approached for multiple POCT on multiple fingersticks. Phase II: n=150 patients approached for multiple POCT on a single fingerstick. The following POC tests were performed by a dedicated nurse: PIMA (CD4), HemoCue (hemoglobin), Reflotron (alanine aminotransferase, creatinine). A venepuncture specimen was taken for predicate laboratory methodology. Normal laboratory ranges and Royal College of Pathologists Australasia (RCPA) allowable differences were used as guidelines for comparison. In 67% of participants, ≥3 tests were requested per visit. All POCT were accurate but ranged in variability. Phase I: Hemoglobin was accurate (3.2%CV) while CD4, alanine aminotransferase and creatinine showed increased variability (16.3%CV; 9.3%CV; 12.9%CV respectively). PIMA generated a misclassification of 12.4%. Phase II: Hemoglobin, alanine aminotransferase and creatinine showed good accuracy (3.2%CV, 8.7%CV, 6.4%CV respectively) with increased variability on CD4 (12.4%CV) but low clinical misclassification (4.1%). No trends were observed for the sequence in which POC was performed on a single fingerstick. Overall, PIMA CD4 generated the highest error rate (16-19%).

Conclusions

Multiple POCT for ART initiation and/or monitoring can be performed practically by a dedicated nurse on multiple fingersticks. The process is as accurate as predicate methodology and can be simplified using a single fingerstick.