Affiliation:1. Faculty of Pharmaceutical Sciences, Tokyo University of Science, 2641 Yamazaki, Noda-shi, Chiba 278-8510, Japan;2. Department of Biochemistry, Keio University School of Medicine, 35 Shinano-machi, Shinjuku-ku, Tokyo 160–8582, Japan;3. Endocrinology & Diabetes Center, Yokohama Rosai Hospital, 3211 Kozukue-cho, Kohoku-ku, Yokohama 222-0036, Japan;4. Department of Uro-Oncology, Saitama Medical University International Medical Center, 1397-1 Yamane, Hidaka-shi, Saitama 350-1298, Japan
Abstract:A method for simultaneous detection of picomole quantities of glutathione (GSH), glutathione disulfide (GSSG), glutathione S-sulfonate (GSSO3H), and cysteine S-sulfonate (CYSSO3H) by high-performance liquid chromatography has been developed. Compounds are separated by anion-exchange chromatography using a citric acid buffer system, and then derivatized postcolumn using o-phthalaldehyde with 2-mercaptoethanol, heated to 70 degrees C, and detected by fluorescence. The compounds elute with retention times of 12.5 min for GSH, 27.5 min for CYSSO3H, 29.8 min for GSSG, and 33.0 minutes for GSSO3H, with detection limits of 10, 200, 10, and 50 pmol, respectively. Recoveries are 103% for GSH, 102% for GSSG, 100% for CYSSO3H, and 96% for GSSO3H. Determination of target compounds in cells is described.