首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Quantitative mass spectrometric immunoassay of insulin like growth factor 1
Authors:Nelson Randall W  Nedelkov Dobrin  Tubbs Kemmons A  Kiernan Urban A
Institution:Intrinsic Bioprobes, Inc, Tempe, Arizona 85281, USA. rnelson@intrinsicbio.com
Abstract:Reported in this work are the development of mass spectrometric immunoassay (MSIA) devices and methods for the qualitative analysis of IGF-1 and -2, and the rigorous quantification of IGF-1 from human plasma. A method involving addition of SDS in moderate concentration to unfractionated plasma for disrupting IGF/IGFBP complexes was initially developed. The method is suitable for the direct extraction of the IGFs and subsequent mass spectrometric analysis. Rat plasma, containing IGF-1 that is mass shifted from human IGF-1, was used as an internal reference standard (IRS) for the quantification of IGF-1 directly from human plasma. A standard curve with linear dynamic range of at least 2 orders of magnitude was constructed from serially diluted IGF-1 standards containing equal amounts of rat plasma. Using the standard curve, IGF-1 levels in plasma samples from eight individuals were determined. The limit of detection for the IGF-1 MSIA was also evaluated and established to be approximately 15 pM. The assay is rapid and can be performed in parallel via high-throughput robotics processing. Furthermore, the mass spectrometry aspect of the developed IGF-1 immunoassay offers a new dimension in the ongoing study of IGF-1 and related diseases.
Keywords:
本文献已被 PubMed 等数据库收录!