A sensitive HPLC-MS-MS method for the determination of raltegravir in human plasma
Authors:Long Mary C  Bennetto-Hood Chantelle  Acosta Edward P
Institution:Division of Clinical Pharmacology, University of Alabama at Birmingham School of Medicine, Birmingham, AL 35294-0019, United States.
Abstract:This work describes an assay system that has been developed to quantify raltegravir concentrations in human plasma using a liquid-liquid extraction technique paired with HPLC separation and MS-MS detection. The dynamic range of this assay extends from 1 to 3000 ng/mL, with a coefficient of determination (r(2), mean+/-SD) of 0.9992+/-0.0002. The mean precision values for calibration standards ranged from 0.6% to 3.0%, while accuracy values were 96.5-104.3%. This procedure is an accurate, precise, and sensitive method for raltegravir quantitation and was successfully validated using external proficiency testing.