Abstract: | This paper describes the development of an isocratic reversed-phase high-performance liquid chromatographic method for the routine analysis of recombinant interleukin-2 (rIL-2) in liposome samples. The chromatographic system employed a C4 column maintained at 30°C eluted with 52.5% (w/w) acetonitrile in water, containing 100 mM NaClO4 and 10 mM HClO4. To remove phospholipid interference the chromatographic method was combined with a lipid-extraction procedure. No significant loss of rIL-2 was noted upon inclusion of this extraction step. The protein eluted from the column with a capacity factor (k′) of 5.8. The method was validated for robustness, linearity, precision and reproducibility. It was shown that the method was linear over a sample concentration range of 1–100 μg/ml. Upon assessment of the intra-day and inter-day precision, the relative standard deviations (RSD) were within the range of the methodical error (approximately 5%), except at the lower concentration of 10 μg/ml, where the intra-day RSD was relatively high (17.8%). The recovery of rIL-2 upon liposome preparation and subsequent analysis of the samples was in the range 94±9%. The results indicate that the method is suitable for routine quantitation of rIL-2 in liposomal samples. |