The BD FACSPresto Point of Care CD4 Test Accurately Enumerates CD4+ T Cell Counts |
| |
Authors: | Priska Bwana Lara Vojnov Maureen Adhiambo Catherine Akinyi Joy Mwende Marta Prescott Matilu Mwau |
| |
Affiliation: | 1. Kenya Medical Research Institute, Nairobi, Kenya.; 2. Clinton Health Access Initiative, Boston, Massachusetts, United States of America.; Rush University, UNITED STATES, |
| |
Abstract: | ObjectiveCurrently 50% of ART eligible patients are not yet receiving life-saving antiretroviral therapy (ART). Financial constraints do not allow most developing countries to adopt a universal test and offer ART strategy. Decentralizing CD4+ T cell testing may, therefore, provide greater access to testing, ART, and better patient management. We evaluated the technical performance of a new point-of-care CD4+ T cell technology, the BD FACSPresto, in a field methods comparison study.Methods264 HIV-positive patients were consecutively enrolled and included in the study. The BD FACSPresto POC CD4+ T cell technology was placed in two rural health care facilities and operated by health care facility staff. We compared paired finger-prick and venous samples using the BD FACSPresto and several existing reference technologies, respectively.ResultsThe BD FACSPresto had a mean bias of 67.29 cells/ul and an r2 of 0.9203 compared to the BD FACSCalibur. At ART eligibility thresholds of 350 and 500 cells/ul, the sensitivity to define treatment eligibility were 81.5% and 77.2% and the specificities were 98.9% and 100%, respectively. Similar results were observed when the BD FACSPresto was compared to the BD FACSCount and Alere Pima. The coefficient of variation (CV) was less than 7% for both the BD FACSCalibur and BD FACSPresto. CD4+ T cell testing by nurses using the BD FACSPresto at rural health care facilities showed high technical similarity to test results generated by laboratory technicians using the BD FACSPresto in a high functioning laboratory.ConclusionsThe BD FACSPresto performed favorably in the laboratory setting compared to the conventional reference standard technologies; however, the lower sensitivities indicated that up to 20% of patients tested in the field in need of treatment would be missed. The BD FACSPresto is a technology that can allow for greater decentralization and wider access to CD4+ T cell testing and ART. |
| |
Keywords: | |
|
|