Abstract: | We measured endotoxin and bacterial levels in tap water, in water purified by reverse osmosis, and in dialysate samples over a 4-month period in a new 10-bed renal dialysis unit. Water treated by reverse osmosis is conducted to the 10 stations through 111 m of piping composed of acrylonitrile butadiene styrene (ABS). All determinations were made prior to the opening of the unit and after the system was purged for 35 h with all bedside station taps open. Formaldehyde disinfection of the piping system was attempted with a recommended protocol after 11 weeks by feeding 2.5 liters of 37% formaldehyde (0.85%, vol/vol) into the delivery system. Prior to water purging, 24 ng of endotoxin per ml was detected. This level decreased to 2.0 ng of endotoxin after the purging. Levels of endotoxin remained below 1.0 ng of endotoxin per ml throughout the duration of the study. In contrast, the level of viable microorganisms recovered from the treated water was approximately 3.5 X 10(4) CFU/100 ml. Even after disinfection of the system, there was no significant decrease in culturable bacteria from the water even though endotoxin levels were lower. Species isolated from the renal dialysis system were predominately pseudomonads, whereas species isolated from the tap water were Bacillus and Flavobacterium species. ABS provides a surface suitable for long-term colonization and growth of bacteria. Currently recommended decontamination protocols are ineffective in removing potentially pathogenic bacteria from ABS pipes and thus constitute an increased risk to patients undergoing dialysis. |