Multi-axial spine biomechanical testing system with speckle displacement instrumentation
Authors:Chung S M  Teoh S H  Tsai K T  Sin K K
Affiliation:Center for Biomedical Materials Applications and Technology, National University of Singapore.
Abstract:This paper reports on the design and development of a multi-axis (up to 6 axes) mechanical tester for spinal research and testing. The developed spine tester allowed true motion to be simulated on a specimen in pure or combined modes. To demonstrate the capability of the new tester flexural stiffness properties of sheep lumbar motion segments were evaluated together wiith a non-contact speckle displacement measurement system. The flexural stiffness of the specimens was measured and compared under constrained and non-constrained testing conditions; with relieving of shear forces (non-constrained), it was found that the specimen behaved in a 'stiffer' manner.