Measurement of cefuroxime in human bronchoalveolar lavage fluid by high-performance liquid chromatography after solid-phase extraction |
| |
Authors: | M.T. Rosseel R. Peleman H. Van Hoorebeke R.A. Pauwels |
| |
Affiliation: | aHeymans Institute of Pharmacology, University of Gent, Medical School, De Pintelaan 185, B-9000 Gent, Belgium;bUniversity Hospital, Department of Respiratory Diseases, De Pintelaan 185, B-9000 Gent, Belgium |
| |
Abstract: | A sensitive and selective method for the determination of cefuroxime in bronchoalveolar lavage (BAL) fluid using high-performance liquid chromatography (HPLC) with UV detection at 280 nm after solid-phase extraction with C18 cartridges was developed. A Waters symmetry C18 column was used and the mobile phase was acetonitrile-0.05 M ammonium phosphate buffer (pH 3.2) (15:85, v/v). The method enabled the determination of cefuroxime at concentrations below 100 ng/ml, with a linear calibration curve at concentrations of 5–100 ng/ml for 400 μl of BAL. The intra- and inter-assay coefficient of variations for 10, 40 and 80 ng/ml were between 5.3 and 8.9%. Analytical recoveries were between 92.7 and 106.2%. The detection limit was 1 ng/ml at a signal-to-noise ratio of 3:1 using 400 μl of BAL. The method was successfully used for the analysis of BAL fluid from patients after oral administration of 500 mg cefuroxime axetil twice daily. |
| |
Keywords: | Cefuroxime |