首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Colorimetric method for the determination of ethanol by Flow Injection Analysis
Authors:AM Salgado  ROM Folly  B Valdman  O Cos  F Valero
Institution:(1) Universitat Autonoma de Barcelona, 08193 Bellaterra, Spain;(2) Departamento de Engenharia Química – Escola de Química –, Universidade Federal do Rio de Janeiro, Rio de Janeiro C.T., B1 E sl 211, Brasil;(3) Departamento d'Enginyeria Química, ETSE, Brasil
Abstract:An automated Flow Injection Analysis system using stop-flow technique for quantifying ethanol based in a colorimetric detection method was developed. The system permitted analysis in a linear range of 0.05–1 g ethanol l–1 without external dilution, a sampling frequency of 15 analyses per hour, and a relative standard deviation of 3.5%. A dilution line was implemented in the FIA system permitting the extension of the linear range to 0.5–5.2 g ethanol l–1 maintaining the same sampling frequency and standard deviation. The system was applied to measure ethanol concentrations present in samples of an alcoholic fermentation and the results showed no significant difference with other analytical procedures (GC).
Keywords:ethanol  fermentation  flow injection analysis  sensor
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏