首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Minimizing tissue-material interaction in microsensor for subcutaneous glucose monitoring
Authors:Ahmad Farook  Christenson Andreas  Bainbridge Martina  Yusof Ahmad Pauzi Mohd  Ab Ghani Sulaiman
Institution:Pusat Pengajian Sains Kimia, Universiti Sains Malaysia, 11800 USM, Pulau Pinang, Malaysia.
Abstract:A new implantable electrocatalytic glucose sensor for subcutaneous glucose monitoring has been fabricated by immobilizing glucose oxidase on a chemically modified carbon fiber. The sensor was inserted subcutaneously on a male spraguely rat without any incision after dipping the microsensor in the rat's serum for 3 days. The so called "stained" microsensor, operated in the amperometric mode with an applied potential of +0.23 V versus Ag|AgCl, was able to directly measure the glucose concentration upon infusion of glucose. The results obtained were encouraging, with the response time was less than 2s and the apparent Michaelis-Menten value at 5.1+/-0.5mM. The "stained" microsensor shows good stability and reproducibility with constant response spanned over 25 days. Most common interferences in glucose analysis were minimized by the outerlayer Nafion. Hematology examinations showed minimal material-tissue interaction. Use of such mechanical devices will allow a more refined understanding towards glucose control in diabetic patients as the implanted microsensor was not effected by biocompatibility failures.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号