A miniaturized planar patch-clamp system for transportable use
Authors:Boussaoud Adrien  Fonteille Isabelle  Collier Guilhem  Kermarrec Frédérique  Vermont Fabien  Tresallet Eric  De Waard Michel  Arnoult Christophe  Picollet-D'hahan Nathalie
Affiliation:CEA, DSV iRTSV, Biologie à Grande Echelle, F-38054 Grenoble, France.
Abstract:In the last decade, planar patch-clamp (PPC) has emerged as an innovative technology allowing parallel recordings of cellular electrophysiological activity on planar substrates. If PPC is widely adopted by the pharmaceutical sector, it remains poorly extended to other areas (i.e. environment and safety organizations) probably because of the large, expensive and non-easily transportable format of those commercial equipments. The present work describes for the first time a new compact and transportable planar patch-clamp system (named Toxint'patch or TIP, for Toxin detection with integrated patch-clamp) focusing on environmental matters and meant to be used in coastal laboratories, for direct on-site monitoring of the seawater and shellfish quality. The TIP system incorporates silicon chips tailored to monitor cellular ionic currents from cultured cells stably expressing a phycotoxin molecular target. The functionality of this novel briefcase-sized PPC system is described in terms of fluidic control, electronic performances with amplifying and filtering boards and of user interface for data acquisition and control implemented on a computer.
Keywords:
本文献已被 PubMed 等数据库收录!