Simultaneous estimation of nicotine and cotinine levels in biological fluids using high-resolution capillary-column gas chromatography combined with solid phase extraction work-up
Authors:Harald W A Teeuwen  Ronald J W Aalders  Jacques M Van Rossum
Institution:(1) Department of Pharmacology Faculty of Medicine, University of Nijmegen, P.O. Box 9101, 6500 HB Nijmegen, The Netherlands;(2) Faculty of Natural Sciences and Mathematics, University of Nijmegen, P.O. Box 9101, 6500 HB Nijmegen, The Netherlands;(3) Netherlands Institute for Drugs and Doping Research (NIDDR), State University of Utrecht, Vondellaan 14, 3521 GE Utrecht, The Netherlands
Abstract:A rapid and sensitive capillary gas-chromatographic method with nitrogen-sensitive detection is reported for the simultaneous analysis of nicotine and cotinine levels occurring in the plasma, saliva, and urine of regular tobacco smokers. The proposed assay has a linear output, has satisfactory accuracy over the range of concentrations of both amines encountered in active smokers, and has also been successful in the analysis of the urine samples of passive smokers. Its lower limit of sensitivity is 0.2 ng of nicotine and 0.5 ng of cotinine per ml of plasma or saliva or per 100 mgrl of urine.The beneficial characteristics of the presented method were achieved by the combination of solid phase extraction of 0.1–1.0 ml of fluid specimens, capillary column gas chromatography with splitless injection and nitrogen sensitive detection, and the use of separate, structurally analogous compounds as internal standards for nicotine. The suitability of the assay is shown by plasma concentration-time curves of nicotine and cotinine in a steady smoker during a 24 hours period.
Keywords:nicotine and cotinine levels  gas-chromatographic method  nitrogen-sensitive detection
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号