Abstract:A new system for continuous measurement of oxygen uptake by means of a telemeter has been developed. Oxygen uptake and pulmonary ventilation during rest and exercise were determined using a portable oxygen consumption meter (Oxylog). A small interface circuit between the Oxylog and the transmitter of a frequency modulated bio-telemeter system was designed and installed inside the Oxylog. Data from the transmitter were passed to a receiver and were fed into a microcomputer system. The microcomputer system displayed and printed out minute values of ventilation and oxygen uptake. The accuracy and reliability of the new system were checked by comparison with the traditional (Douglas bag) method. In the range less than 80 l.min-1 of ventilation and less than 2 l.min-1 of oxygen uptake, the system was not inferior to the Douglas bag method. The new system was applied for field continuous measurement of oxygen uptake during a doubles tennis game. The results of the application indicate that the telemetry system developed here is a very practical and useful way of measuring oxygen uptake during sports activities.