Colorimetric determination of hydroxyurea in human serum using high-performance liquid chromatography
Authors:Konstantine K Manouilov  Timothy R McGuire  Peter R Gwilt  
Abstract:A high-performance liquid chromatography assay for hydroxyurea in human serum was developed based on a commercial colorimetric assay kit for urea (Sigma Diagnostics). Serum (0.5 ml), spiked with methylurea as an internal standard, was treated with 70% perchloric acid. Supernatant (0.2 ml) was combined with 0.7 ml of BUN acid reagent and 0.6 ml of BUN color reagent. The resulting colored reactant (100 μl) was analyzed on a 300×3.9 mm Bondclone 10 C18 column coupled with a UV–Vis detector, at 449 nm. The mobile phase was 13% acetonitrile in water. Retention times of colored derivatives of hydroxyurea and methylurea were 6.5 and 12.2 min, respectively. The log–log calibration curve was linear from 0.0065 to 1.31 mM. Average accuracy was 99.9±4.0% and the intra- and inter-day error of assay did not exceed 11%.
Keywords:Hydroxyurea