首页 | 本学科首页   官方微博 | 高级检索  
     


The role of H2O2 outer diffusion on the performance of implantable glucose sensors
Authors:Vaddiraju S  Burgess D J  Jain F C  Papadimitrakopoulos F
Affiliation:Nanomaterials Optoelectronics Laboratory, Polymer Program, Institute of Materials Science, University of Connecticut, Storrs, CT 06269, United States.
Abstract:The performance of an implantable glucose sensor is strongly dependent on the ability of their outer membrane to govern the diffusion of the various participating species. In this contribution, using a series of layer-by-layer (LBL) assembled outer membranes, the role of outwards of H(2)O(2) diffusion through the outer membrane of glucose sensors has been correlated to sensor sensitivity. Glucose sensors with highly permeable humic acids/ferric cations (HAs/Fe(3+)) outer membranes displayed a combination of lower sensitivities and better linearities when compared with sensors coated with lesser permeable outer membranes (namely HAs/poly(diallyldimethylammonium chloride) (PDDA) and poly(styrene sulfonate) (PSS)/PDDA). On the basis of a comprehensive evaluation of the oxygen dependence of these sensors in conjunction with the permeability of H(2)O(2) through these membranes, it was concluded that the outer diffusion of H(2)O(2) is crucial to attain optimized sensor performance. This finding has important implications to the design of various bio-sensing elements employing perm-selective membranes.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号