首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Electrical detection of human immunoglobulins G from human serum using a microbiosensor
Authors:Marcon Lionel  Stiévenard Didier  Melnyk Oleg
Institution:Institut d'Electronique de Microélectronique et de Nanotechnologie, UMR CNRS 8520, Dpt ISEN, 41 Bd Vauban, 59046 Lille Cedex, France.
Abstract:A biosensor for the electrical detection of human antibodies from serum has been fabricated and experimentally demonstrated. The device is based on the immobilization of proteins used as probes between a set of microelectrodes. Incubation with diluted human serum was followed by incubation with anti-human secondary antibodies labeled with gold nanoparticles (GNPs) and then precipitation of silver on the nanoparticles. The output of the device was defined as the percentage of short-circuited microelectrodes after silver deposition independently of the gap conductance. Two model probes were studied: protein A and goat antibodies. The effects of the microgap spacing (5, 10, 15 or 20 microm) and the duration of the silver treatment were examined. The data obtained showed that a large spacing (20 microm) led to poor sensitivity. Alternately, 5 microm gaps led to high sensitivity and saturation of the signal. Interestingly, 10-15 microm gaps enabled a non-saturated and distinct signal for both probes that was correlated with the GNP density between the microgaps as determined by atomic force microscopy. Different capture efficiencies could be easily distinguished. The biosensor described here is easy to use and thus can be applied to real detection experiments.
Keywords:
本文献已被 PubMed 等数据库收录!