High-performance liquid chromatographic assay with ultraviolet detection for the determination of etoperidone and two active metabolites, 5-(1-hydroxyethyl)etoperidone and 1-(3-chlorophenyl)piperazine, in plasma |
| |
Authors: | Mary L. Holland Edward T. Heebner |
| |
Affiliation: | Mary L. Holland*,Edward T. Heebner |
| |
Abstract: | A selective and sensitive high-performance liquid chromatographic assay with ultraviolet detection for the determination of the antidepressant drug etoperidone and two active metabolites in plasma is described. The drug, metabolites and internal standard are isolated from plasma using a two-step liquid—liquid extraction procedure. The resulting sample is chromatographed on a C18 column (10 cm × 2.1 mm I.D.) with ultraviolet detection at 254 nm. Standard curves are linear for each compound over the concentration range 2–1000 ng/ml. The accuracy and precision of the assay, expressed as the percentage deviation of measured values from the true value and the relative standard deviation (inter-run), are ≤ 10% at all concentrations except the minimum quantification limit. Using an automated injector and computerized data acquisition, eighty samples can be routinely processed in one day. The assay has been successfully used for the analysis of plasma samples from pharmacokinetic studies in mice, rats, dogs and humans. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|