Determination of trace lead in biological and water samples with dispersive liquid-liquid microextraction preconcentration
Authors:Liang Pei  Sang Hongbo
Institution:aKey Laboratory of Pesticide & Chemical Biology of Ministry of Education, College of Chemistry, Central China Normal University, Wuhan 430079, People’s Republic of China
Abstract:A new method for the determination of trace lead was developed by dispersive liquid–liquid microextraction preconcentration and graphite furnace atomic absorption spectrometry. In the proposed approach, 1-phenyl-3-methyl-4-benzoyl-5-pyrazolone (PMBP) was used as a chelating agent, and carbon tetrachloride and ethanol were selected as extraction and dispersive solvents. Some factors influencing the extraction efficiency of lead and its subsequent determination, including extraction and dispersive solvent type and volume, pH of sample solution, concentration of the chelating agent, and extraction time, were studied and optimized. Under the optimum conditions, the enrichment factor of this method for lead was reached at 78. The detection limit for lead was 39 ng L−1 (3σ), and the relative standard deviation (RSD) was 3.2% (n = 7, c = 10 ng mL−1). The method was successfully applied to the determination of trace amounts of lead in human urine and water samples.
Keywords:Dispersive liquid–  liquid microextraction  Lead  Graphite furnace atomic absorption spectrometry  Biological and water samples