首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Capillary electrophoretic separation of proteins and peptides by ion-pairing with heptanesulfonic acid
Authors:Miksík I  Charvátová J  Eckhardt A  Cserháti T  Forgács E  Deyl Z
Institution:Institute of Physiology, Academy of Sciences of the Czech Republic, Vídenská 1083, CZ-14220 Prague 4, Czech Republic. miksik@biomed.cas.cz
Abstract:Heptanesulfonic acid as ion-pairing agent was used for the separation of mixtures of low and high molecular mass peptides/proteins by capillary electrophoresis. The separation conditions used were: capillary 37 cm (30 cm to the detector) x 75 microm i.d., voltage 10 kV, phosphate buffer 50 mmol/l, ion-pairing agent heptanesulfonic acid at three different concentrations, namely, 0, 20 or 100 mmol/l, pH 2.5. The separation reflected the ion-pairing equilibria between the ion-pairing agent and the peptide/protein analytes. The influence of ion-pairing on sample mobility (running time) was more pronounced in case of the higher-molecular peptides as compared to the low molecular ones. This difference offers the possibility to separate low and high molecular peptides/proteins that under the absence of the ion-pairing agent would co-migrate. The principle of this approach was demonstrated on a randomly selected set of peptides/proteins; the practical applicability was demonstrated on a set of CNBr peptides arising from a naturally occurring mixture of collagen types I and III.
Keywords:
本文献已被 PubMed 等数据库收录!