首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Comparison of the QIAsymphony automated nucleic acid extraction and PCR setup platforms with NucliSens easyMAG and Corbett CAS-1200 liquid handling station for the detection of enteric pathogens in fecal samples
Authors:van Zanten E  Wisselink G J  de Boer W  Stoll S  Alvarez R  Kooistra-Smid A M D
Institution:
  • a Laboratory for Infectious Diseases, Department of Research and Development, van Ketwich Verschuurlaan 92, 9721SW, Groningen, The Netherlands
  • b Qiagen, Qiagenstr 1, D-40724, Hilden, Germany
  • Abstract:In this study we compared the QIAsymphony Sample Preparation and Assay Setup modules (Qiagen), which provide automated nucleic acid extraction and PCR setup, to the NucliSens easyMAG (bioMérieux) and CAS-1200 liquid handling station1 (Corbett) for a molecular screening approach of enteric pathogens in fecal samples using multiplex real-time PCR. The relative DNA recovery of both platforms, within- and between-run reproducibility and a prospective study, including 510 clinical fecal samples, were performed. The results demonstrated that the QIAsymphony Sample Preparation and Assay Setup modules were highly reproducible and achieve equal performance, quantitative and qualitative, when compared with the NucliSens easyMAG and CAS-1200 systems for the molecular screening analysis of enteric pathogens in fecal samples.
    Keywords:QIAsymphony SP/AS  NucliSens easyMAG  CAS-1200/QIAgility  Real-time PCR  Enteric pathogens
    本文献已被 ScienceDirect PubMed 等数据库收录!
    设为首页 | 免责声明 | 关于勤云 | 加入收藏

    Copyright©北京勤云科技发展有限公司  京ICP备09084417号