首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Automated large-volume sample stacking procedure to detect labeled peptides at picomolar concentration using capillary electrophoresis and laser-induced fluorescence detection
Authors:Siri Nathalie  Riolet Pierre  Bayle Christophe  Couderc François
Institution:Picometrics, 10 Avenue de l'Europe, 31520 Ramonville, France.
Abstract:We have developed an automated large-volume sample stacking (LVSS) procedure to detect fluorescein isothiocyanate-labeled peptides in the picomolar range. The injection duration is 10 min at 50 mbar to fill 62% of the capillary volume to the detection cell. The calculated limit of detection (S/N=3), filling 1% of the capillary volume, is 74 pM for bradykinin and 45 pM for L-enkephalin with samples diluted in water and analyzed in a 50 mM borate buffer, pH 9.2. With the automated LVSS system, the limits of detection are 7 pM for bradykinin, 3 pM for L-enkephalin and 2 pM for substance P. LVSS is shown to be quantitative from 500 to 10 pM.
Keywords:Large-volume sample stacking  Peptides  Fluorescein isothiocyanate
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏