The Self-Expanding Symetis Acurate Does Not Increase Cerebral Microembolic Load When Compared to the Balloon-Expandable Edwards Sapien Prosthesis: A Transcranial Doppler Study in Patients Undergoing Transapical Aortic Valve Implantation |
| |
Authors: | Gabor Erdoes Christoph Huber Reto Basciani Stefan Stortecky Stephan Windecker Peter Wenaweser Thierry Carrel Balthasar Eberle |
| |
Affiliation: | 1. Department of Anesthesiology and Pain Therapy, Inselspital, University Hospital Bern, Bern, Switzerland.; 2. Department of Cardiovascular Surgery, Inselspital, University Hospital Bern, Bern, Switzerland.; 3. Department of Cardiology, Inselspital, University Hospital Bern, Bern, Switzerland.; Colorado State University, United States of America, |
| |
Abstract: | ObjectivesThe aim of this study was to quantify potential differences in count, frequency and pattern of high-intensity transient signals (HITS) during transapical transcatheter aortic valve implantation (TA-TAVI), by comparing the Symetis Acurate TA (SA) with the balloon-expandable Edwards Sapien XT (ES) system.BackgroundRecently, the Symetis Acurate TA revalving system has been introduced for TA-TAVI. The Symetis Acurate TA aortic bioprosthesis is self-expanding and is deployed by a specific two-step implantation technique. Whether this novel method increases the load of intraprocedural emboli, detected by transcranial Doppler ultrasound (TCD) as HITS, or not is not clear.MethodsTwenty-two patients (n = 11 in each study arm, median logistic EuroScore 20%, median STS score 7%) displayed continuous TCD signals of good quality throughout the entire TA-TAVI procedure and were included in the final analysis. Data are presented as median with interquartile ranges.ResultsNo significant differences were detected in total procedural or interval-related HITS load (SA: 303 [200; 594], ES: 499 [285; 941]; p = 0.16). With both devices, HITS peaked during prosthesis deployment (PD), whereas significantly fewer HITS occurred during instrumentation (SA: p = 0.002; ES: <0.001) or post-implantation PI (SA: p = 0.007; ES: <0.001). PD-associated HITS amounted to almost half of the total HITS load. One patient suffered new disabling stroke at 30 days. Thirty-day mortality amounted to 13.6% (3 of 22 patients).ConclusionsSimplified transapical delivery using the self-expanding SA device does not increase HITS, despite of a two-step deployment technique with more interactions with the native aortic valve, when compared to the balloon-expandable ES valve. The similarity in HITS count, frequency and pattern with the two systems suggests a common mechanism for the release of cerebral microemboli. |
| |
Keywords: | |
|
|