The Self-Expanding Symetis Acurate Does Not Increase Cerebral Microembolic Load When Compared to the Balloon-Expandable Edwards Sapien Prosthesis: A Transcranial Doppler Study in Patients Undergoing Transapical Aortic Valve Implantation
Authors:Gabor Erdoes  Christoph Huber  Reto Basciani  Stefan Stortecky  Stephan Windecker  Peter Wenaweser  Thierry Carrel  Balthasar Eberle
Affiliation:1. Department of Anesthesiology and Pain Therapy, Inselspital, University Hospital Bern, Bern, Switzerland.; 2. Department of Cardiovascular Surgery, Inselspital, University Hospital Bern, Bern, Switzerland.; 3. Department of Cardiology, Inselspital, University Hospital Bern, Bern, Switzerland.; Colorado State University, United States of America,
Abstract:

Objectives

The aim of this study was to quantify potential differences in count, frequency and pattern of high-intensity transient signals (HITS) during transapical transcatheter aortic valve implantation (TA-TAVI), by comparing the Symetis Acurate TA (SA) with the balloon-expandable Edwards Sapien XT (ES) system.

Background

Recently, the Symetis Acurate TA revalving system has been introduced for TA-TAVI. The Symetis Acurate TA aortic bioprosthesis is self-expanding and is deployed by a specific two-step implantation technique. Whether this novel method increases the load of intraprocedural emboli, detected by transcranial Doppler ultrasound (TCD) as HITS, or not is not clear.

Methods

Twenty-two patients (n = 11 in each study arm, median logistic EuroScore 20%, median STS score 7%) displayed continuous TCD signals of good quality throughout the entire TA-TAVI procedure and were included in the final analysis. Data are presented as median with interquartile ranges.

Results

No significant differences were detected in total procedural or interval-related HITS load (SA: 303 [200; 594], ES: 499 [285; 941]; p = 0.16). With both devices, HITS peaked during prosthesis deployment (PD), whereas significantly fewer HITS occurred during instrumentation (SA: p = 0.002; ES: <0.001) or post-implantation PI (SA: p = 0.007; ES: <0.001). PD-associated HITS amounted to almost half of the total HITS load. One patient suffered new disabling stroke at 30 days. Thirty-day mortality amounted to 13.6% (3 of 22 patients).

Conclusions

Simplified transapical delivery using the self-expanding SA device does not increase HITS, despite of a two-step deployment technique with more interactions with the native aortic valve, when compared to the balloon-expandable ES valve. The similarity in HITS count, frequency and pattern with the two systems suggests a common mechanism for the release of cerebral microemboli.