Development of a Peristaltic Micropump for Bio-Medical Applications Based on Mini LIPCA
Authors:Thanh Tung Nguyen    My Pham  Nam Seo Goo
Institution:[1]Department of Aerospace Engineering, Konkuk University, Seoul 143-701, South Korea [2]Department of Advanced Technology Fusion, Konkuk University, Seoul 143-701, South Korea [3]Biomemitics and Intelligent Microsystem Laborotary, Artificial Muscle Research Center, Konkuk University, Seoul 143-701, South Korea
Abstract:This paper presents the design, fabrication, and experimental characterization of a peristaltic micropump. The micropump is composed of two layers fabricated from Polydimethylsiloxane (PDMS) material. The first layer has a rectangular channel and two valve seals. Three rectangular mini lightweight piezo-composite actuators are integrated in the second layer, and used as actuation parts.Two layers are bonded, and covered by two Polymethyl Methacrylate (PMMA) plates, which help increase the stiffness of the micropump.A maximum flow rate of 900 uL·min-1 and a maximum backpressure of 1.8 kPa are recorded when water is used as pump liquid. We measured the power consumption of the micropump. The micropump is found to be a prom- ising candidate for bio-medical application due to its bio-compatibility, portability, bidirectionality, and simple effective design.
Keywords:peristaltic micropump  LIPCA  PDMS  bio-medical applications