Repeatability Using Automatic Tracing with Canon OCT- HS100 and Zeiss Cirrus HD-OCT 5000
Authors:Rune Brautaset  Ulrika Birkeldh  Petra Frehr Alstig  Petra Wikén  Maria Nilsson
Institution:Unit of Optometry, Department of Clinical Neuroscience, St. Erik Eye Hospital, Karolinska Institutet, Stockholm, Sweden;Stanford University Medical Center, UNITED STATES
Abstract:

Background

Optical coherence tomography (OCT), can be used in clinical practice to provide high resolution cross-sectional images of the retina, optic disc and macula structure. These measurements can be useful for early detection, diagnosis, monitoring and treatment guidance for retinal diseases. Therefore, repeatability of measurements in OCT is of great importance.

Methods

Macula and optic disc parameters from the right eye of 30 healthy subjects were obtained twice with the Canon OCT-HS100 and Zeiss Cirrus HD-OCT 5000. Repeatability was evaluated by use of the coefficient of repeatability (CR) and the coefficient of repeatability as a percentage of the mean (CR%), and the obtained measurements were compared between the instruments.

Results

CR% of optic disc parameters ranged between 0.90 and 22.22% and 0.00 and 16.00% with the Canon and Zeiss OCT respectively. For macular parameters CR% ranged between 0.62 and 2.81% and 0.99 and 1.81% with the Canon and Zeiss OCT respectively. No statistical difference could be found when comparing the CR of all macular and disc measurements between the instruments. Compared to our previously published data repeatability has significantly improved with the inclusion of automatic tracking systems with both the Canon and Zeiss OCT.

Conclusion

Automatic tracking function improves repeatability in both Canon OCT-HS100 and Zeiss Cirrus HD-OCT 5000. However, measurements generated by the two instruments are still not interchangeable.